DEPOLOX
®
5 E flow cell
Retrofit kits, spares and accessories
6.
59
10
Insert the optional sensor measuring module cards in the provided
slots. See chapter 6.2.3 “DFMe electronics module”.
11
When installing the SiDiSens LF module and/or the total chlorine
membrane sensor TC3 CAN, switch off the CAN-balancing resistors
on the DFMe electronics module. Set the DIP switch (G - Figure 3) to
„OFF” position. Furthermore, the CAN extension socket (I - Figure 2)
must also be retrofitted. To do this, break out the indent on the hous
-
ing underside of DFMe electronics module and fit the CAN extension
socket (I - Figure 2). Check that all gasket inserts are correctly
mounted.
12
Fit the housing cover of DFMe electronics module and secure with
the four cover screws. Tighten the housing screws to a maximum
torque of 0.7 Nm (± 0.15 Nm).
13
Hook DFMe electronics module by the mounting hook (D - Figure 2)
onto DEPOLOX
®
5 E flow cell and secure on the underside with the
securing screws.
Image 2 Cross-section, DFMe electronics module
Image 3 Cutout of circuit board
A Multi-sensor
B Sensor cable Cl
2
free
C LED glow stick
D Mounting hook
E pH sensor module
F Redox sensor module
G DIP switch
H Motherboard DFMe with Measurement input Cl
2
I CAN extension socket
J CAN connection socket to 700 P electronics module or 700 M
electronics module
A
B
C
E
F
D
I
J
G
H
D
I
G
Содержание WALLACE & TIERNAN DEPOLOX 5 E 700 P Series
Страница 1: ...WALLACE TIERNAN DEPOLOX 5 E FLOW CELL INSTRUCTION MANUAL...
Страница 2: ...2 WT 050 810 040 DE IM 0519 DEPOLOX 5 E flow cell Please note Original instruction manual Simone...
Страница 56: ...56 WT 050 810 040 DE IM 0519 Maintenance DEPOLOX 5 E flow cell 5...
Страница 72: ...72 WT 050 810 040 DE IM 0519 Declarations and certificates DEPOLOX 5 E flow cell 7...
Страница 76: ...76 WT 050 810 040 DE IM 0519 Declarations and certificates DEPOLOX 5 E flow cell 7...
Страница 80: ...80 WT 050 810 040 DE IM 0519 Index DEPOLOX 5 E flow cell 8...