60
WT.050.810.040.DE.IM.0519
Retrofit kits, spares and accessories
DEPOLOX
®
5 E flow cell
6.
14
Install the SiDiSens LF module on the left side in the flow cell and
secure with the enclosed plastic self-tapping screws (W2T807965 -
10 mm) in position N - Figure 4 (mounting recess).
Please note
Prior to the installation of the SiDiSens module, please ensure that you
use the correct plastic self-tapping screws. Use only the plastic self-tap
-
ping screws A2 (W2T807965 - 10 mm) when installing the SiDiSens
module. Please distinguish between the plastic self-tapping screws for
installing the flow cell and those used for the SiDiSens module. The
screws should not be confused.
15
Plug in the CAN sensor cable of the SiDiSens LF module at
position L - Figure 4. The M12 terminating resistor must be screwed
on CAN socket (O - Figure 5).
Image 4 DEPOLOX
®
5 E flow cell, non-pressurized (cover removed)
K CAN sensor cable connection
L SiDiSens LF connection
M Membrane sensor connection
N Mounting recess
Plastic self-tapping
screw (10 mm)
K
M
N
L
Содержание WALLACE & TIERNAN DEPOLOX 5 E 700 P Series
Страница 1: ...WALLACE TIERNAN DEPOLOX 5 E FLOW CELL INSTRUCTION MANUAL...
Страница 2: ...2 WT 050 810 040 DE IM 0519 DEPOLOX 5 E flow cell Please note Original instruction manual Simone...
Страница 56: ...56 WT 050 810 040 DE IM 0519 Maintenance DEPOLOX 5 E flow cell 5...
Страница 72: ...72 WT 050 810 040 DE IM 0519 Declarations and certificates DEPOLOX 5 E flow cell 7...
Страница 76: ...76 WT 050 810 040 DE IM 0519 Declarations and certificates DEPOLOX 5 E flow cell 7...
Страница 80: ...80 WT 050 810 040 DE IM 0519 Index DEPOLOX 5 E flow cell 8...