DEPOLOX
®
5 E flow cell
Description
3.
23
Redox combined measur
-
ing and reference elec
-
trode
(W3T169298 – platinum
version)
Conductivity sensor
(W3T172052)
Design
Combined measuring and reference elec
-
trode with platinum electrode, salt reserve,
zirconium dioxide diaphragm, polymerized
solid electrolyte, Ag/AgCl drain system
Measurement range
±2000 mV
Working temperature
range
-5 to +80°C (23 to 176°F)
Operating pressure
0 to 6 bar (6 × 10
5
Pa)
Minimum conductivity
of the sample water
50 µS/cm
Installation length
120 mm
Screw-in thread
PG 13.5
Storage temperature
-5 to +30°C (23 to 86°F)
Design
4-electrode measurement, integrated tem
-
perature sensor NTC 30, graphite elec
-
trodes, epoxy shaft
Measurement range
1 µS/cm to 2 S/cm
Cell constant
0.475 cm
-1
±1.5%
Working temperature
range
-5 to +100°C (23 to 212°F)
Operating pressure
(electrode shaft)
0 to 10 bar (1 × 10
6
Pa)
Installation length
120 mm
Installation
Loose
Protection rating
IP 67 (in plugged in state)
Storage temperature
0 to +50°C (32 to 122°F), store in the fresh
air
Содержание WALLACE & TIERNAN DEPOLOX 5 E 700 P Series
Страница 1: ...WALLACE TIERNAN DEPOLOX 5 E FLOW CELL INSTRUCTION MANUAL...
Страница 2: ...2 WT 050 810 040 DE IM 0519 DEPOLOX 5 E flow cell Please note Original instruction manual Simone...
Страница 56: ...56 WT 050 810 040 DE IM 0519 Maintenance DEPOLOX 5 E flow cell 5...
Страница 72: ...72 WT 050 810 040 DE IM 0519 Declarations and certificates DEPOLOX 5 E flow cell 7...
Страница 76: ...76 WT 050 810 040 DE IM 0519 Declarations and certificates DEPOLOX 5 E flow cell 7...
Страница 80: ...80 WT 050 810 040 DE IM 0519 Index DEPOLOX 5 E flow cell 8...