DEPOLOX
®
5 E flow cell
Installation
4.
45
4.4.14
Fitting the sens (optional)
Please note
The membrane sensors, combined measuring and reference electrodes
and electrodes must be prepared accordingly. Please follow the appro
-
priate instructions for the sensors!
The pressurized version can be retrofitted with the total chlorine mem
-
brane sensor TC3 CAN. Please note that the standard cell body cover
must be replaced by an optional cell body cover, part no. W3T417351.
Depending on requirements, insert or screw in the sensors into the mount
hole on the cover of the cell body.
Please note
With the pressurized version, the sensors must be screwed in or secured
to prevent them from being pushed out.
Proceed as follows:
1
On the non-pressurized version, remove the cap from the mounting
hole on the cell body cover.
2
On the pressurized version, unscrew the protection cap.
3
Take the pH and redox combined measuring and reference electrode
out of the KCl tank with stand.
Please note
Keep the KCl tank with stand for later use or storage.
4
With the conductivity sensor, remove the blind plug and O-ring.
Содержание WALLACE & TIERNAN DEPOLOX 5 E 700 P Series
Страница 1: ...WALLACE TIERNAN DEPOLOX 5 E FLOW CELL INSTRUCTION MANUAL...
Страница 2: ...2 WT 050 810 040 DE IM 0519 DEPOLOX 5 E flow cell Please note Original instruction manual Simone...
Страница 56: ...56 WT 050 810 040 DE IM 0519 Maintenance DEPOLOX 5 E flow cell 5...
Страница 72: ...72 WT 050 810 040 DE IM 0519 Declarations and certificates DEPOLOX 5 E flow cell 7...
Страница 76: ...76 WT 050 810 040 DE IM 0519 Declarations and certificates DEPOLOX 5 E flow cell 7...
Страница 80: ...80 WT 050 810 040 DE IM 0519 Index DEPOLOX 5 E flow cell 8...