WT.050.560.001.UA.IM.0714
103
Depolox
®
3 plus ReSIDUAl ANAlYZeR
eVoQUA
W3T140188
(1) Counter electrode (blue spot) blue wire
(4) earth grey wire
(2) Working electrode (red spot) pink wire
(5) Temperature sensor brown wire
(3) Reference electrode white wire
(6) Temperature sensor yellow & green wire
Figure 4.1
m. Reassemble the cover (10) and cable gland (11) to the electrode hous-
ing.
n. Refit the electrode/electrolyte reservoir assembly and secure with the
clamping nut (2).
o. Reconnect the water inlet line, outlet line and the cell drain line.
p. Leave the cell running for an hour, to allow the system to settle down,
and then carry out the zero calibration procedure. This procedure may
be repeated after 24 hours if required.