Signal Diagnosis Menu (Cont’d)
Display
Value Range
(defaults bold)
Description
REL: 1 2 3 4
St: 1 0 0 0
Informations about the relay status
1: alarm relay 1 (0: unpowered/1: powered)
2: alarm relay 2
3: alarm relay 3
4: alarm relay 4
Press Enter and then the up or down arrow to select the
relay, to change the relay state for testing press Enter.
After leaving this menu the relay goes back to previous
state.
DI: 1 2 3
St: 1 0 0
Informations about the digital status
1: digital input 1 0: closed (inactive)
2: digital input 2 1: open (active)
3: digital input 3
VF Conv. AE1
14.07 kHz
Actual measurement of the input 1
(disinfection)
VF Conv. AE2
14.07 kHz
Actual measurement of the input 2
(pH or fluoride)
A/D (PT1000) AE3
356 Digit
or
A/D (PT100) AE3
356 Digit
Actual measurement of the input 3
(temperature)