WT.050.560.001.UA.IM.0714
74
Depolox
®
3 plus ReSIDUAl ANAlYZeR
eVoQUA
W3T140188
If the sensor’s measuring signal is still too low or irregular, a new membrane
cap must be used. The measuring cell requires one to two hours before it
can be calibrated.
3.4.6.4
Membrane Sensor for Total Chlorine TC1
The membrane sensor measures the total amount of free and combined chlo-
rine (chloramine). The sensor incorporates a pH buffer in the electrolyte and
is suitable for water with various pH values. The sensor may only be used in
clear water of potable water quality.
CAUTION: The electrode fingers (reference electrodes) and membrane are
extremely sensitive! Do not touch, soil or damage! Before unscrewing the
filled membrane cap, push the elastomer seal to one side to permit the inflow
of air through the vent underneath it! Otherwise, the membrane may be
damaged due to developing negative pressure. Do not sand down the light
yellowish-gray deposit of the electrode finger or wipe it in the direction of
the gold working electrode!
Commissioning the Membrane sensor for Total Chlorine TC1
1. Screw off the membrane cap from the electrode shaft and fill to the
top of the threads on inside of cap with the included gel electrolyte.
2. Rub gold working electrode with the included lapping paper (special
emery). To do this, lay the lapping paper on a paper towel, take hold
of the corner and while holding the measuring cell vertically, slide the
electrode end one to two times over the rough side of the lapping paper.
3. Check whether the elastomer seal completely closes the valve opening.
4. Then place the electrode shaft on the filled membrane cap and slowly
screw the membrane cap in place by hand.
NOTE: Because excess electrolyte escapes through the hose valve, do
not clamp it shut and do not press on the elastomer seal.
5. Rinse away the escaped electrolyte with water.
!
Содержание Wallace & Tiernan Depolox 3 plus
Страница 11: ...WT 050 560 001 UA IM 0714 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 12: ......
Страница 19: ......
Страница 39: ...WT 050 560 001 UA IM 0714 26 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 40: ......
Страница 109: ......
Страница 127: ...WT 050 560 001 UA IM 0714 112 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 128: ......
Страница 146: ...WT 050 560 001 UA IM 0714 130 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 147: ......
Страница 157: ......