![Evoqua Wallace & Tiernan Depolox 3 plus Скачать руководство пользователя страница 102](http://html1.mh-extra.com/html/evoqua/wallace-and-tiernan-depolox-3-plus/wallace-and-tiernan-depolox-3-plus_manual_2447569102.webp)
WT.050.560.001.UA.IM.0714
88
Depolox
®
3 plus ReSIDUAl ANAlYZeR
eVoQUA
W3T140188
Key
Abbreviations of the data
formats
Bool-boolean, UCHAR-unsigned char, SCHAR-signed char, USINT-un-
signed integer, SINT-signed integer, UloNG-unsigned long, SloNG-signed
long, FloAT- float, ASCII-ASCII-code, MIx-mixed format
Abbreviations of access
l-read, S-write, lp-read with password, Sp-write with password
Data structure 1
Byte1-2: Measured valuet (signed int), Byte3-4: range-start (signed int),
Byte 5-6: range-end (signed int), Byte7-11: unit (ASCII), Byte12: divider
(character)
Table 3.4 - Address Reference List
ADDR.
DeSCRIpTIoN
FoRMAT leNGTH
RANGe
UNIT FACToR STATUS
DeSCRIpTIoN
BYTeS
0.
Date of software version
ASCII
12
-
-
-
l
-
0
1.
Module name
ASCII
28
-
-
-
l
-
12
2.
password of interface
USINT
2
0...999
-
1.0
l, S
-
40
3.
Module type
ASCII
12
1
-
-
l
-
42
4.
Measurement value DeS
USCHAR
12
Meas. range mg/l
0.10
0.01
l
Data Structure 1
54
5.
Measurement value
pH/Fluoride
USCHAR
12
Meas. range pH
mg/l
0.01
l
Data Structure 1
66
6.
Measurement value
temperature
USCHAR
12
Meas. range
°
C
0.10
l
Data Structure 1
78
7.
errors
USloNG
4
-
-
-
l
0x00000001: Digital input 1
0x00000002: Digital input 2
0x00000004: Digital input 3
0x00000008: Calibration error DeS
0x00000010: Calibration error pH
0x00000020: Calibration error Fluor
0x00000040: Cell error
0x00000080: Cell error pH
0x00000100: Cell error Fluor
0x00000200: Temperature
0x00000400: load mA output
0x00000800: ADU1
0x00001000: ADU2
0x00002000: ADU3
0x00004000: Values out of range
0x00008000: Max DeS
0x00010000: Min DeS
0x00020000: Max pH
0x00040000: Min pH
0x00080000: Max Fluor
0x00100000: Min Fluor
0x00200000: System
0x00400000: oVR DeS
0x00800000: oVR pH
0x01000000: oVR Fluor
0x02000000: HoCl pH
0x04000000: --
0x08000000: --
0x10000000: --
0x20000000: --
0x40000000: --
0x80000000: --
90
8.
Cell current DeS
FloAT
4
-
µA
1.0
l
-
94
9.
Cell voltage pH/Fluor
FloAT
4
-
mV
1.0
l
-
98
10.
State of leDs and DI and
relays
USINT
2
-
-
-
l
0x0001: leD Alarm 1
0x0002: leD Alarm 2
0x0004: leD Alarm 3
0x0008: leD Alarm 4
0x0010: leD DoS
0x0020: leD DI
0x0040: leD Failure
0x0080: DI1
0x0100: DI2
0x0200: DI3
0x0400: Relay K1
0x0800: Relay K2
0x1000: Relay K3
0x2000: Relay K4
102
11.
Menu language
USCHAR
1
-
-
-
l, Sp
0x01: German
0x02: english
0x04: French
0x08: Spanish
104
12.
Function mA output
USCHAR
1
-
-
-
l, Sp
0x01: off
0x02: 0-20mA
0x04: 4-20mA
105
Содержание Wallace & Tiernan Depolox 3 plus
Страница 11: ...WT 050 560 001 UA IM 0714 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 12: ......
Страница 19: ......
Страница 39: ...WT 050 560 001 UA IM 0714 26 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 40: ......
Страница 109: ......
Страница 127: ...WT 050 560 001 UA IM 0714 112 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 128: ......
Страница 146: ...WT 050 560 001 UA IM 0714 130 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 147: ......
Страница 157: ......