WT.050.560.001.UA.IM.0714
10
Depolox
®
3 plus ReSIDUAl ANAlYZeR
eVoQUA
W3T140188
4 5 6 7 8 9 10
oFF
2.1.4 Select Sensor Types
The setting for the measuring signal of the disinfection sensors must correspond
to the connected sensor. Select the sensor at the ten-position DIP switch S3
on the board (factory setting is for Bare Electrode).
WARNING: DO NOT CHANGE SWITCH SETTINGS UNTIL UNIT IS DISCON-
NECTED FROM LINE POWER. CHANGING THE SWITCH SETTINGS WHILE
THE UNIT IS POWERED UP WILL DAMAGE THE ELECTRONICS AND LOCK
UP THE SySTEM PROGRAM, AND VOID SySTEM WARRANTy.
WARNING: WRONG SETTING CAN CAUSE DAMAGE TO THE SENSOR AND
THE DEPOLOx® 3
plus
ELECTRONICS.
Figure 2.2 - DIP Switch
NOTE: Disconnect line power before changing DIP switch settings. Failure to
disconnect power will damage electronics and void warranty.
3-Elect. “Depolox® 3
plus
”
(Bare Electrode):
(Default factory setting)
!
!
switch
Содержание Wallace & Tiernan Depolox 3 plus
Страница 11: ...WT 050 560 001 UA IM 0714 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 12: ......
Страница 19: ......
Страница 39: ...WT 050 560 001 UA IM 0714 26 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 40: ......
Страница 109: ......
Страница 127: ...WT 050 560 001 UA IM 0714 112 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 128: ......
Страница 146: ...WT 050 560 001 UA IM 0714 130 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 147: ......
Страница 157: ......