WT.050.560.001.UA.IM.0714
46
Depolox
®
3 plus ReSIDUAl ANAlYZeR
eVoQUA
W3T140188
3.1.6
Alarm Relays
Depolox® 3
plus
features the following alarm relays:
• Two relays in single electronics unit.
• Four relays in dual electronics unit.
Definitions for switching the relays:
DIS max:
relay reacts to the set value
DIS min:
relay reacts to the set value
pH max:
relay reacts to the set value
pH min:
relay reacts to the set value
Fluor max:
relay reacts to the set value
Fluor min:
relay reacts to the set value
DI1, DI2, DI3:
relay is activated as long as there is a signal on the
digital input
general fault:
relay is activated as long as there is an alarm mes-
sage
dosing contact:
relay is energized as long as the actual value is
below the setpoint (for relay 1 only)
Selections:
normally open (N.O.):
relay is energized in case of an alarm.
fail safe (N.C.):
relay is de-energized in case of an alarm
non latching:
relay is de-energized as soon as the alarm
condition has ceased
latching:
relay stays energized after the alarm until the
star key “
” is pressed to acknowledge the
alarm
3.1.7
Dosing Contact for Chlorine
The chlorine alarm relay 1 can be used to switch-on a pump or solenoid valve
for dosing chlorine.
The relay is energized if “AR 1 Definition” is set to “dosing contact” and
the actual value is < setpoint.
3.1.8
mA Outputs
For each sensor there is an mA current signal output (e.g., for connecting an
additional display). The range fixed at 4...20mA corresponds to the measuring
range. For example, for a Membrane range of 0 to 5 mg/l:
0 mg/l: 4 mA
5 mg/l: 20 mA
Содержание Wallace & Tiernan Depolox 3 plus
Страница 11: ...WT 050 560 001 UA IM 0714 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 12: ......
Страница 19: ......
Страница 39: ...WT 050 560 001 UA IM 0714 26 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 40: ......
Страница 109: ......
Страница 127: ...WT 050 560 001 UA IM 0714 112 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 128: ......
Страница 146: ...WT 050 560 001 UA IM 0714 130 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 147: ......
Страница 157: ......