![Evoqua Wallace & Tiernan Depolox 3 plus Скачать руководство пользователя страница 68](http://html1.mh-extra.com/html/evoqua/wallace-and-tiernan-depolox-3-plus/wallace-and-tiernan-depolox-3-plus_manual_2447569068.webp)
WT.050.560.001.UA.IM.0714
54
Depolox
®
3 plus ReSIDUAl ANAlYZeR
eVoQUA
W3T140188
3.3
Free Chlorine (W2T9694 Bare Electrode) Sensor Kit
3.3.1
Start-Up of the Measuring Cell
a. Unscrew nut (2) and remove cover (3) (see Dwg. 50.560.000.010 in Section
5).
b. Hold electrode assembly together with electrolyte reservoir (4-11) at the
cover (10) and pull it out of the cell block (13) without turning.
c. Remove felt ring (transport protection of the membrane, 6)
d. Exchange plug (4) with separately supplied plug fitted with membrane for
venting the electrolyte reservoir.
e. Place electrode assembly with electrolyte reservoir back into the cell block.
Keying pin in the electrode assembly must snap into position.
f. Fill a half capful of grit into the cell block.
g. Place cover (3) and tighten by nut (2).
h. Allow sample water flow by opening the inlet valve(s).
i. After one hour the measuring signal has stabilized so far that a first calibra-
tion should be made as function test.
NOTE: For the first starting-up: After 24 hours of running time a further cali-
bration is necessary. Air bubbles building up at the inner surface of the cell
block and preventing somewhat the rotation of the grit will disappear after
a running-in period of 24 to 48 hours.
3.3.2 Adjusting the Flow Regulator
The flow regulator is adjusted to a flow of 33 l/h (=0.55 l/min.). If necessary
adjust the regulator as follows:
NOTE: Units are factory calibrated. Adjustment is only required if maintenance
is performed on the flow valve.
a. Remove the left or right hinge pin at the measuring cell and tilt the housing
front to the side.
b. At the inner side of the front is the flow regulator. In the middle of the
regulator is the adjusting screw. Turn the screw with a coin or a screwdriver
(>10 mm). Turn to the right to reduce flow; turn to the left to increase flow.
c. Check flow and reassemble.
Содержание Wallace & Tiernan Depolox 3 plus
Страница 11: ...WT 050 560 001 UA IM 0714 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 12: ......
Страница 19: ......
Страница 39: ...WT 050 560 001 UA IM 0714 26 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 40: ......
Страница 109: ......
Страница 127: ...WT 050 560 001 UA IM 0714 112 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 128: ......
Страница 146: ...WT 050 560 001 UA IM 0714 130 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 147: ......
Страница 157: ......