Depolox
®
3 plus ReSIDUAl ANAlYZeR
eVoQUA
W3T140188
c. Allow water to drain from the flow block.
d. Remove one sensor after the other from the flow block.
e. Clean the sensors and store them as described below.
f. Cover the holes in the cover of the flow block with the protective caps.
To re-start the system, refer to paragraph 2.3, Membrane Sensor.
• To store the sensors after use:
Remove the elastomer seal from the vent hole. Unscrew the membrane
cap. Rinse the membrane and the reference electrode with distilled water.
Carefully dry the reference electrode with a paper towel. Let the cap dry
in a dust free place. Screw the dry membrane cap loosely onto the sensor.
The membrane must not touch the gold working electrode.
3.4.3
VariaSens™ Flow Block Assembly
The VariaSens™ flow block assembly contains the following:
• Cell body with cover
• Flow control valve
• Multi-sensor
• Drain
• Fine filter (For use only when membrane sensors are used.)
• Sample water inlet with check valve and ball valve
Two clips are installed in the housing cover. These are used to facilitate pH and
fluoride calibration. Insert these clips into the rear panel of the housing. Insert
one clip into the upper catch to attach the sensor. Insert the second clip into
the center catch inorder to measure the buffer or calibration solution using
the plastic cups provided.
Содержание Wallace & Tiernan Depolox 3 plus
Страница 11: ...WT 050 560 001 UA IM 0714 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 12: ......
Страница 19: ......
Страница 39: ...WT 050 560 001 UA IM 0714 26 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 40: ......
Страница 109: ......
Страница 127: ...WT 050 560 001 UA IM 0714 112 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 128: ......
Страница 146: ...WT 050 560 001 UA IM 0714 130 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 147: ......
Страница 157: ......