WT.050.560.001.UA.IM.0714
97
Depolox
®
3 plus ReSIDUAl ANAlYZeR
eVoQUA
W3T140188
Table 4.1 - errors (Cont’d)
Error Message
Cause
Remedy
Additional Errors
Device has no display
No mains power
Defective fuse or wrong setting of
mains voltage
Turn on external mains switch
Check voltage setting and replace
fuse, see paragraph 2.1.3
Displayed/output value wrong
Wrong calibration
Old electrolyte or clogged mem-
brane
Wrong wiring or setting
New calibration
Sensor maintenance
Check sensor, wiring and setting
Device not responding correctly to
software adjustments or correc-
tions.
Incorrect software programming
or corrupted memory.
Initialize software, see paragraph
3.1.9.
In case of an alarm:
• the alarm LED will become illuminated
• the display will detail the specific alarm every five seconds for one second
for each alarm.
• if an alarm relay is associated to the alarm condition, the delay time will
begin. At the end of the delay time the relay will be triggered.
NOTE: Refer to Table 4.2 - Relay / Display Operation During Alarm Conditions,
for details.
Содержание Wallace & Tiernan Depolox 3 plus
Страница 11: ...WT 050 560 001 UA IM 0714 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 12: ......
Страница 19: ......
Страница 39: ...WT 050 560 001 UA IM 0714 26 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 40: ......
Страница 109: ......
Страница 127: ...WT 050 560 001 UA IM 0714 112 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 128: ......
Страница 146: ...WT 050 560 001 UA IM 0714 130 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Страница 147: ......
Страница 157: ......