Alarm
Warnings shown on the WaterLily
Vacuum Generator TS display
Green LED
Yellow/
red LED
Speaker
Action to take
POSSIBLE FULL
CANISTER ALARM!
activated if
the fluid collection
canister is full
The alarm page disappears when the problem
is solved
or
after pressing
OK
, the operating page appears
cyclically every 15 sec showing the alarm
silenced symbol and the alarm present symbol
in red
and for 5 seconds
If the alarm condition has not been cleared
2
minutes
after pressing
OK
, the alarm page
reappears permanently
On (device
operational)
Blinking Red
LED on
On
Silenceable for
2
minutes
by
pressing
OK
WARNING
During the alarm
silence time, keep
the patient and
the device
monitored to
check that the
alarm condition
has actually been
cleared.
WARNING
If the canister is full, before replacing it and
resuming the therapy, check that the wound is not
haemorrhaging (check for any signs of bleeding in
the suction tube and the canister). If so, urgent and
appropriate medical action needs to be taken.
Check that the canister has not tilted/tipped over. If so, move it back
into vertical position.
If the alarm automatically resets, the canister does not have to be
replaced, otherwise, even if the canister is not full, replace it and
resume the therapy.
The alarm can be silenced for 2 minutes by pressing
OK
This alarm is automatically reset only when the problem has been
solved.
WARNING
The absence of vacuum in the dressing might
determine build-up of exudates triggering infectious
and inflammatory processes. If the absence of
vacuum extends for several days, necrotic eschar
may form.
If suction is interrupted for more than 2 hours,
remove the dressing after irrigating it with saline
solution.