102
Alarm
Warnings shown on the WaterLily
Vacuum Generator TS display
Green LED
Yellow/
red LED
Speaker
Action to take
EXHAUSTED
BATTERY!
CONNECT THE
BATTERY CHARGER
activated if
the rechargeable
internal battery is
exhausted
The alarm page disappears when the power
supply/battery charger is connected
On (device
operational)
Blinking Red
LED on
On
NOT SILENCEABLE
Power the device directly from the mains using the power
supply/battery charger which will at the same time start a new
recharging cycle.
Information signals
The
WaterLily Vacuum Generator TS
device is equipped with the following visual and acoustic information signals:
Information signal
Warnings shown on the
WaterLily Vacuum Generator
TS display
Green LED Yellow/red
LED
Speaker
Action to take
SEAL TEST
activated
at the start of therapy
until the set suction
pressure has been
reached
On (device
operational)
Off
On
a two-tone
acoustic signal is
sounded when
the set suction
pressure has been
reached.
None.
REMINDER
activated
when the set reminder
time has run out
On (device
operational)
Off
On
a two-tone
acoustic signal is
sounded when
the time has run
out
Reset the reminder signal by pressing
OK
.