76
Basic programming of the suction pressure
Repeatedly press
or
depending on whether you want to increase or decrease the suction
pressure. Each time an arrow is pressed, the suction pressure decreases or increases by 5 mmHg.
Once you have set the new pressure value, a page appears asking you to confirm:
Press
YES
to confirm or
NO
to cancel.
The last suction pressure set remains in memory even if the device is switched off (it switches on again in
CONTINUE THERAPY
mode) and/or the battery goes completely flat.
Replacing the WaterLily TS Canister
To replace the single-use canister when it is full, disconnect it from the wound suction tube/drain and
remove it from the vacuum generator by raising the hook and partially turning it anticlockwise and then
extracting it upwards.
Switching off the device
Press and hold down the
button until the device switches off.
While the device is switching off, you will hear an acoustic signal (beep) and the following will appear on the
display:
Switching on the device and activating CONTINUE THERAPY
Note
When activating the CONTINUE THERAPY function, the preset parameters will
be maintained and chronological recording of the events continued.
Press and hold the
button for a few seconds until the home page followed by the start page appears: