105
•
Limited solely to the power supply/battery charger provided:
-
The power supply/battery charger has passed all the applicable tests according to EN 60601-1
(2006) except those indicated in clause 15.5.3. Referring to clause 15.5.3, the tests passed are
those required by CEIEN 60601-1 (1998) indicated in clause 57.9.
Electromagnetic compatibility
ELECTROMAGNETIC EMISSIONS
WATERLILY VACUUM GENERATOR TS is intended for use in the electromagnetic environment specified
below. The customer
or
the user of WATERLILY VACUUM GENERATOR TS should assure that it is used in
such an environment.
EMISSION TEST
IEC 60601-1-2
TEST LEVEL
ELECTROMAGNETIC ENVIRONMENT GUIDELINES
RF emissions
CISPR 11
Group 1
WATERLILY VACUUM GENERATOR TS only uses RF
energy for its internal function.
Therefore, its RF
emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
WATERLILY VACUUM GENERATOR TS is suitable for
use in all establishments, including domestic
establishments and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Compliant
Warnings
WATERLILY VACUUM GENERATOR TS is in compliance with IEC EN 60601-1-2 on
electromagnetic compatibility for medical electrical equipment.
However, as a
precaution, the device should not be used in the immediate vicinity of high-power
equipment or equipment that, by their nature, emit strong electromagnetic fields. Mobile
phones, portable phones or other radio communications equipment used near the device
might affect its operation.
Should the device necessarily need to be used in the vicinity of other equipment, it
should be observed to check that it is operating normally in the configuration in which it
is used.
Do not use extension cables or adapters on the power cable.
The device must be connected to the mains only with the 4m-long power cable provided;
the device has been tested with this cable and resulted compliant to the Electromagnetic
Compatibility standards. Using cables of different length may cause increased emissions or
reduced immunity against radio disturbances.