Aquafilter
E9_Next_Ing_rev03 – 10/10/2018
81
E9 NEXT
W85
The steriliser is not detecting
the SD memory card, or the
SD card is write-protected.
Check that the SD memory card is present
and correctly installed. Switch off the
steriliser, remove the SD memory card and
check that the safety switch allows writing.
The conductivity read by the
conductivity meter on the
autoclave is at the limits of
the acceptable values.
Empty the clean water tank as soon as
possible, and refill it with demineralised or
distilled water of the best quality.
The conductivity read by the
conductivity meter on the
autoclave is at the limits of
the acceptable values.
Empty the clean water tank, and refill it with
demineralised or distilled water of the best
quality.
W43
Door
lock
electromagnet
faulty.
Switch the steriliser off and then back on
again. If the problem persists contact the
assistance service.
W44
Door lock electromagnet out
with door open.
Manually move the electromagnet back in by
pushing the pin toward the steriliser.
W80
Lower
band
heater
temperature not suitable for
cycle start.
Resistance fault.
W81
Upper band heater
temperature not suitable for
cycle start.
Resistance fault.
W82
Upper band heater
temperature not suitable for
NGV cycle start.
Resistance fault.
W90
Steam generator
temperature sensor faulty.
Switch the steriliser off and then back on
again. If the problem persists contact the
assistance service.
W91
Upper band temperature
sensor faulty.
Switch the steriliser off and then back on
again. If the problem persists contact the
assistance service.
W92
Lower band temperature
sensor faulty.
Switch the steriliser off and then back on
again. If the problem persists contact the
assistance service.
W93
Faulty condensation battery
sensor.
Switch the steriliser off and then back on
again. If the problem persists contact the
assistance service.
W94
Pressure sensor faulty.
Switch the steriliser off and then back on
again. If the problem persists contact the
assistance service.
W95
Chamber temperature
sensor faulty.
Switch the steriliser off and then back on
again. If the problem persists contact the
assistance service.