Aquafilter
E9_Next_Ing_rev03 – 10/10/2018
26
E9 NEXT
Emptying used water
If the
used water tank
is full, the following icon appears on the LCD display of the control panel:
In these cases,
it is not possible to perform sterilisation cycles
. The capacity of the clean water tank is
sufficient for about 8 cycles for the 18L model, and for about 7 cycles for the 24L model.
1. Empty the internal used water collection tank:
CAUTION: DANGER OF CONTAMINATION.
The used water in the discharge tank may contain
contaminated residues: it is advisable to wear latex safety gloves when draining (chap. 3.4
“Residual risks”).
NEVER REUTILISE USED WATER.
Get hold of an empty tank, fit the clear tube supplied with the unit, after removing the discharge section
cover (3 of Fig. 6.8-2) to the connector with the grey button at the bottom of the front panel (2 of Fig. 6.8-
2). At the end of the draining operation, remove the tube from the connector by pressing on the clip.
CAUTION:
this operation is fundamental for the correct operation of the unit.
Fig. 6.8-2
Used water can easily be continuously drained by means of the draining connector situated on the rear part
of the unit (3 of Fig. 6.8-3). After connecting the tube (Fig. 6.8-4), make sure it never exceeds the height of
the connector on the steriliser during its journey to the drain, otherwise the water will not flow (Fig. 6.8-5).