Unloading and preserving sterilized instruments
The material is at the greatest risk of contamination while it is still hot, because the barrier capabilities of the
packaging materials are much lower in the presence of residual humidity, compared to an ambient
temperature situation. The materials, therefore, should not be stacked once they have been extracted, in
order to favour the dispersion of heat.
Wait for the material to drop to room temperature before stocking it: before stocking, make sure that the
packages are intact and check the chemical colour change; if the package is broken or torn, the load can
only be used immediately, in that preservation of sterility cannot be guaranteed.
The material should be stocked in sealed cabinets, 30 cm away from the floor and 5 cm from the ceiling; if
this is not possible, protect the material in nylon bags.
For the duration of the sterile condition refer to the current regulations of the country of use, and
to the statements of the manufacturer of the packaging used.