36
E9 NEXT
7.3.1 Start-up, execution and end of a test
CAUTION: the VACUUM TEST can only be activated with the machine cold, i.e. WITHIN 3
MINUTES FROM SWITCHING ON THE UNIT
, in that once this time has elapsed, the unit starts
pre-heating (see “Appendix 6”). It will no longer be possible to perform the test.
If the unit is pre-heating and you switch it off and then switch it back on again, it still will not be possible to
perform the vacuum test in that the steriliser must be cold.
If, however, the test is completed with positive results, the following screen appears:
At this point, the door release symbol indicates that the door can be opened, and the display goes back to
the Test Selection screen.