Aquafilter
E9_Next_Ing_rev03 – 10/10/2018
62
E9 NEXT
9.3
EXTRAORDINARY MAINTENANCE
Any jobs not mentioned above are considered as extraordinary maintenance. In these cases, contact
specialists authorised by Euronda S.p.A.
CAUTION:
extraordinary maintenance must be performed by qualified staff only.
The bacteriological filter and the gasket are components that are not covered by the
guarantee.
Service maintenance
After 1000 cycles or after two years from installation it is necessary to carry out a general overhaul of the
unit. This can only be performed by specialists authorised by Euronda S.p.A.
ATTENTION: extraordinary maintenance must only be performed by specialists authorised
by Euronda S.p.A..
Replacing the bacteriological filter
-
Unscrew the bacteriological filter (7 of Fig. 5.1.1-2) by turning it anticlockwise
-
Screw on the new filter by turning it clockwise until it is tight.
Replacing the door seal
-
Grip the lip of the seal with two fingers and remove it;
-
Clean the seat of the seal with a cloth soaked in alcohol;
-
Fit the new seal into the seat located in the door and distribute it evenly around the circumference by
applying the same pressure on the entire gasket with your fingers. Then lift up the lip of the gasket to
make sure no points have been badly fitted.
-
Switch on the steriliser, close the door making sure the correct closing force is required; if necessary,
adjust the closing force with the relative adjustment wrench.
To simplify insertion of the new gasket in the slot use neutral soap diluted 1 to 10 with
water.
Cleaning the drain filter
When necessary, clean the drain filter at the front of the boiler; unscrew the filter as shown in the figure (fig.
9.3-1), rinse it under running water and then screw back in place.
Fig. 9.3-1