Aquafilter
E9_Next_Ing_rev03 – 10/10/2018
30
E9 NEXT
7.2
SELECTING A STERILIZATION CYCLE
To select a sterilisation system, press one of the five available cycle buttons on the CYCLES screen:
After the cycle has been selected, this screen appears:
This screen summarises the main characteristics of the chosen cycle:
•
at the top, the cycle name (B 134 in this case);
•
next to the cube image, the maximum admissible solid load (4.5/6 kg in this case);
•
next to the wave image, the maximum admissible porous load (1.5/2 kg in this case);
•
next to the zig-zag line, the number of pre-vacuum cycles for removing the air required by the cycle (3 in
this case);
•
next to the flat line, the sterilisation time (4 minutes in this case);
•
next to the hour glass, the average cycle duration (48 minutes in this case).
At the bottom, the green hand for running the cycle and the arrow-left icon for returning to the previous
screen.
For loads not exceeding 0.6 kg, of solid and 0.2 kg of porous, placed on the same tray, it is possible to
perform a rapid cycle allowing the load to be sterilised in an average time of 30 minutes. The RAPID cycle
has a shorter drying time that dries the load in any case, even if it is placed in bags.