If the Shortcut Call is enabled, the digital 0~9 keys will change the
original function, only the calling function is retained, and the
display status bar will display SHORTCUT. When Shortcut Call is
turned off, the numeric keys will return to normal.
4.2.3 Trigger Output Settings
The E-load has a TRO signal output port at the rear panel. In the auto mode and
over-current protection mode, when an output signal is required, you can program the
output signal, output condition, and the action after the test failed.
For example: set the output mode to Level, the output condition to Qualified and failed
action to Stop.
The TRO port level will change from high to low when the test is qualified, and will remain
the status until the next test begins. If a failure is encountered, the output is always high
and the test is stopped.
For more specific settings, refer to
Chapter 4.2 Parameter Settings - Output
and
Chapter 6.1 I/O Port Settings
.
4.3 Save
/
Recall Function
The E-load can save parameters and working-mode up to 10 groups for convenient and
fast usage. For example, first set load parameters in CC mode, set the range under
MENU, and go to the SAVE for storage, the SAVE is as follows:
Figure 4.7 Save Setting Page
ET System electronic GmbH