26
Users can edit up to 8 sets of list files. If the E-load operation mode is List operation, the
E-load will start a List operation when the key ON/OFF is pressed till completion or the
ON/OFF is pressed again to stop the loading.
Figure 3.7 LIST loading
For example:
Source under test:
constant voltage source 24V, maximum output current 5A
Purposes:
to test current output and voltage fluctuation at 5 voltage points of 1A, 2A, 3A,
4A, and 5A.
Operation steps:
1. Press the key
to enter the mode selection page, press the up and down keys to
select LIST, and press the key
to enter the LIST test interface.
2. In Figure 3.5, press the key
to enter the setting interface, as shown in Figure 3.6.
3. Select the desired file number by pressing the key + /-, as shown in Figure 3.6
4. Use the up, down, left and right direction keys to select the mode as CONT (the
counting function only valid in CNT mode).
5. Set the first step current to 1A,time to 1mS and current rate to 3A/us through the up,
down, left and right keys.
6. Press the key ADD to increase the number of steps and set the parameters for each
step, as shown in Figure 3.6.
7. After the setting is completed, press the key
to back to the list test interface,
ET System electronic GmbH