37
Pulse
In pulse-mode, with dynamic test operation enabled, the
load will switch to Ib every time after receipt of a trigger
signal and switch back to Ia value after maintaining Tb for
pulse width time.
Toggle
In toggle mode, the load current will drop to Ia according to
the set current rate, or rise to Ib according to the set current
rate after receipt of every trigger signal.
DYNA setting parameters
Parameter
Description
Ia
Low level loading current
Ta
Low level current duration, setting range 10uS~50S
Ib
High level loading current
Tb
High level current duration, setting range 10uS~50S
Rise Rate
Current rise time rate A/uS
Fall Rate
Current fall time rate A/uS
Mode
Continuous/ Pulse/ Toggle
Continuous mode
In continuous mode and dynamic function enabled, the E-load continuously switch
between two load values
Figure 3.18 DYNA- Continuous mode
Pulse mode
In pulse mode, after enabling dynamic test operation, the E-load will switch to Ib value