40
Figure 3.22 Battery mode setting page
BATT Setting parameters
Parameters
Description
Mode
Discharge mode CC/CW/CR
Value
Load value in different discharge modes
Stop Time
Discharge stop time (termination conditions)
Unit
AH/WH
Stop Volt
Discharge stop voltage (termination conditions)
Stop Cap
Discharge stop capacity (termination conditions)
After the above setting confirmed, press the key
to start testing. When the set stop
conditions is reached, the test will be stopped and the load input status turn to be OFF.
Press the key
to stop testing, the
light wil be OFF and the load input status turns
to be OFF, the battery stops discharging.
Source under test:
18650 battery rated voltage 3.7V, capacity 2400mAh.
Purpose:
check the battery with a capacity of 2400 mAh when it discharging from
full-charge to the lowest voltage
ET System electronic GmbH