Iii. Select the baud rate you want. At present, the E-load only supports 9600, 19200,
38400 and 57600.
8.4 SCPI Communication Instruction
1
*IDN?
Meaning: Query version number
Return:8151, V1.0
Example: query version number
Send: *IDN?
Back: 8151,V1.0
2
*TRG
Meaning: Trigger test, no return
Example: Trigger test
Send: *TRG
Back: no
3
INPut
Meaning: input on or off, 0 off 1 on
Example: Input on
Send: INPut 1
Back: no
4
CURRent:RANGe
Meaning: set current range LOW/HIGH 0/1
Example: set low current range
Send: CURRent: RANGe 0
Back: no
5
VOLTage:RANGe
Meaning: set the voltage range LOW/HIGH(0/1)
Example: set the voltage range to LOW
Send: VOLTage: RANGe 0
Back: no
6
CURRent:SLEW
Meaning: set current rise time and fall time
Example: set the current rise/fall time to 3A/uS
Send: CURRent: SLEW 3
Back: no
7
CURRent:SLEW:RISE