28
Figure 3.10 OCP parameter setting interface
OCP parameter
For example:
Scource under test: constant voltage source 24V, maximum output current 5A
Purposes: to judge the current value of over-current protection point between 4.8A and
5.2A or not.
Operation steps:
1.
Press the key
to enter the mode selection page, press the up and down keys to
select OCP, and press the key
to enter the OCP test interface.
2.
In Figure 3.8, press the key
to enter the setting interface, as shown in Figure
3.10. Use the up, down, left and right direction keys to set the parameters as shown in
Figure 3.10.
OCP parameter:
Istart
Start Current
Iend
End Current
Step
Steps
(
1~1000
)
Dwell
Step duration
(
0.01~999.99
)
V trig
Trigger voltage level
Ocp Lo
Lower limit of over-current
Ocp Hi
Upper limit of over-current
Compare
Comparison model
ET System electronic GmbH