Troubleshooting
96
Platinous User’s Manual
The viewing window is
clouded or frosted over.
Humidity rose sharply.
There is nothing wrong with the
equipment. Proceed as planned.
The fuse in the frost prevention
heater blew.
Replace fuse F14.
If the new fuse
blows, call for service.
The chamber is wet on the
outside.
The room is highly humid.
There is nothing wrong with the
equipment. Proceed as planned.
Frost has formed around
the door hinges, door and
viewing window (ESX,ESZ)
The room is highly humid.
There is nothing wrong with the
equipment. Proceed as planned.
The fuse in the frost prevention
heater blew.
Replace fuse F14.
If the new fuse
blows, call for service.
Temperature (& Humidity) is
unstable.
The door is open.
Close the door.
The rubber plug fell off the cable.
Fit the plug into the port.
Ambient temperature changes more
than 5°C/hr.
Stabilize ambient temperature and
resume testing.
High heat load equipment is being
turned ON/OFF.
Reduce the heat load.
The water supply filter is clogged.
Clean the filter.
Temperature gradually rises
higher than the target tem-
perature.
Specimens are generating heat.
Reduce the amount of heat gener-
ated by specimens.
Frost has formed on the cooler.
Defrost the chamber.
Temperature rises or low-
ers too slowly.
The door is open.
Close the door.
Specimens are generating heat.
Reduce the amount of heat gener-
ated by specimens.
Ambient temperature is too low or
too high.
Raise or lower the ambient tempera-
ture.
Temperature stops rising or
begins to drop during heat-
up.
Frost has formed on the dehumidi-
fier-cooler.
There is nothing wrong with the
equipment. Proceed as planned. Or,
defrost the chamber.
Poor Temperature unifor-
mity.
Air flow inside the chamber is poor.
Improve air flow.
Specimens are generating heat.
Reduce the amount of heat gener-
ated by specimens.
Frost has formed on the cooler.
The chamber lamp does not
light up.
Fuse F13 has blown.
Replace fuse F13.
If the new fuse
blows, call for service.
The fluorescent bulb is blown.
See “Changing the Chamber Lamp
Bulb” on page 97.
The starter is blown.
See “Changing the Chamber Lamp
Starter” on page 98.
TABLE 8.18.
O
PERATOR
ERROR
AND
ADDITIONAL
TROUBLESHOOTING
Trouble
Cause
Remedial Action
Содержание Platinous ESL Series
Страница 1: ...PLATINOUS CHAMBER SERIES USER S MANUAL...
Страница 2: ......
Страница 6: ...iv Platinous User s Manual...
Страница 22: ...Precautions in Usage 16 Platinous User s Manual...
Страница 31: ...Platinous User s Manual 25 Names and Functions of Parts...
Страница 40: ...Names and Functions of Parts 34 Platinous User s Manual...
Страница 58: ...Installation 52 Platinous User s Manual...
Страница 80: ...Operation 74 Platinous User s Manual...
Страница 106: ...Troubleshooting 100 Platinous User s Manual...
Страница 107: ...Platinous User s Manual Chapter 9 Appendix...