Installation
42
Platinous User’s Manual
5.5 Cooling Water Connection (Water-Cooled
chambers only)
This option uses a water-cooled condenser in the refrigeration system.
Equipment
•
Condenser: Water-cooled type.
•
Condenser Fan: The condenser is not equipped with a fan.
•
Water Supply and Drain Ports: Refrigerator water supply and drain ports are located
on the chamber’s left side.
Chamber Performance
Chamber performance is guaranteed for a water temperature of 25°C, no specimens and automatic
refrigeration capacity control.
Water Supply and Drainage Work
Water-cooled chambers require water supply and drainage lines. First, check water supply rate and
pressure to satisfy the levels below:
TABLE 5.11.
W
ATER
SUPPLY
RATE
AND
PRESSURE
Water Supply Rate
(gpm)
Water Supply
Pressure
Water Temperature (°F)
(Working Range)
ESL
3 @ 75°F
MInimum 30 psi
Pressure Differential
Maximum 100 psig
10 ~ 85
6 @ 85°F
ESU
3 @ 75°F
6 @ 85°F
ESX
4 @ 75°F
8 @ 85°F
ESZ
4 @ 75°F
8 @ 85°F
Содержание Platinous ESL Series
Страница 1: ...PLATINOUS CHAMBER SERIES USER S MANUAL...
Страница 2: ......
Страница 6: ...iv Platinous User s Manual...
Страница 22: ...Precautions in Usage 16 Platinous User s Manual...
Страница 31: ...Platinous User s Manual 25 Names and Functions of Parts...
Страница 40: ...Names and Functions of Parts 34 Platinous User s Manual...
Страница 58: ...Installation 52 Platinous User s Manual...
Страница 80: ...Operation 74 Platinous User s Manual...
Страница 106: ...Troubleshooting 100 Platinous User s Manual...
Страница 107: ...Platinous User s Manual Chapter 9 Appendix...