Installation
38
Platinous User’s Manual
5.2 How to Install
DO NOT move the chamber with specimens inside.
Specimens could overturn or fall off shelves during transport.
Install the chamber only on a flat, level floor.
An unleveled floor could trigger an alarm or destabilize humidity control.
Always anchor the chamber in place with the adjuster feet BEFORE use.
Unless anchored, the chamber could move unexpectedly under external force.
The chamber must be level side-to-side and front-to-back for proper humidity system operation.
Casters and adjuster feet are found under each of the four corners of the chamber base.
Adjuster foot height can be adjusted by loosening and tightening the nut. This will require a 1/2” or 13mm
open-end wrench.
FIGURE 5.20.
H
OW
TO
LOOSEN
/
TIGHTEN
ADJUSTER
FEET
To Level the Chamber
1. Lower the adjuster feet until the casters rise off the floor.
2. Once installed, check that the chamber is level from side to side and front to back, using a level placed
on the top of the chamber.
3. Remove the shipping bracket from the door.
To Move the Chamber
1. Install the shipping bracket on the door.
2. Raise the adjuster feet until the chamber is sitting on the casters.
3. Move the chamber.
!
CAUTION
Содержание Platinous ESL Series
Страница 1: ...PLATINOUS CHAMBER SERIES USER S MANUAL...
Страница 2: ......
Страница 6: ...iv Platinous User s Manual...
Страница 22: ...Precautions in Usage 16 Platinous User s Manual...
Страница 31: ...Platinous User s Manual 25 Names and Functions of Parts...
Страница 40: ...Names and Functions of Parts 34 Platinous User s Manual...
Страница 58: ...Installation 52 Platinous User s Manual...
Страница 80: ...Operation 74 Platinous User s Manual...
Страница 106: ...Troubleshooting 100 Platinous User s Manual...
Страница 107: ...Platinous User s Manual Chapter 9 Appendix...