Platinous User’s Manual
93
Troubleshooting
Displayed Alarm
Code/Channel
Code (option)
Trouble
Cause
Remedial Action
HUMIDIFIER
WATER SUPPLY
RATE2
(26)
The humidifying tray did
not fill within the specified
amount time during
humidity testing. The
chamber has been
stopped. If the backup
mode is ON, humidity
tests are suspended, but
temperature tests
continue.
The strainer on
the humidifying
water supply
circuit is either
clogged or the
circuit is leaking,
or water supply
pressure is too
low.
Check the humidifying water
supply circuit (wait for the test
end if the backup mode is on)
and correct as necessary.
Then, resume testing. If the
same alarm occurs again, call
for service.
DRY WICK
(26)
During humidity testing,
the wet-bulb (measure
relative humidity) rose
above the specified
temperature. The
humidifier will continue as
is, but humidity control
might destabilize and
trigger a humidity alarm.
The wet-bulb
wick (woven
cloth) inside the
chamber may be
dry.
Stop the chamber and replace
the wick. After wet-bulb
temperature returns within the
specified range, the alarm can
be cleared by pressing the
Buzzer Stop key.
COMMUNICATION
ERROR-1
(99)
The instrumentation
detected communication
trouble between the CPU
board and display board.
The chamber has been
stopped.
Communication
error between the
CPU board and
display board
Reactivate the system from
the main power switch
(leakage breaker) and resume
testing. If the same alarm
occurs again, call for service.
COMMUNICAITON
ERROR-2
(99)
The instrumentation
detected system trouble
in the display board. The
chamber has been
stopped.
Display board
error (i.e. internal
memory, etc.)
Reactivate the system from
the main power switch
(leakage breaker) and resume
testing. If the same alarm
occurs again, call for service.
COMMUNICATION
ERROR-4
(99)
The instrumentation
detected system trouble
in the CPU board. The
chamber has been
stopped.
Sequence task
error on CPU
board.
Reactivate the system from
the main power switch
(leakage breaker) and resume
testing. If the same alarm
occurs again, call for service.
Содержание Platinous ESL Series
Страница 1: ...PLATINOUS CHAMBER SERIES USER S MANUAL...
Страница 2: ......
Страница 6: ...iv Platinous User s Manual...
Страница 22: ...Precautions in Usage 16 Platinous User s Manual...
Страница 31: ...Platinous User s Manual 25 Names and Functions of Parts...
Страница 40: ...Names and Functions of Parts 34 Platinous User s Manual...
Страница 58: ...Installation 52 Platinous User s Manual...
Страница 80: ...Operation 74 Platinous User s Manual...
Страница 106: ...Troubleshooting 100 Platinous User s Manual...
Страница 107: ...Platinous User s Manual Chapter 9 Appendix...