When running in the constant mode OR when running in the program mode but not needing to
continue after clearing the alarm
Procedure:
1. Press the BUZZER STOP button on the Alarm screen to silence the buzzer.
2. If the chamber is running, press either the OPER./START key on the operating panel or the chamber
operating status box on the screen to get the Operation Mode Selection screen. Then press the STOP
button under Stop Operation. Note the alarm name(s) before switching power OFF.
3. Press the POWER key to deactivate control power.
4. Set the main power breaker in the OFF position.
5. Remedy the trouble for the alarm displayed on the Alarm screen as explained in the alarm table.
6. Set the main power breaker in the ON position and then activate control power. The display will come
on and the main menu will appear shortly.