Platinous User’s Manual
73
Operation
Specifications
•
Number of outputs: 1
•
Contact rating: 125V AC, 2A
Operation
A signal is output via a closed contact when the chamber has been stopped by a safety device tripping.
The interface contact for the alarm signal closes when an alarm condition on the chamber occurs.
Usage
Connect the alarm circuit (i.e. alarm emitter, etc.) to the external alarm output on the Chamber Options
Panel. The alarm circuit must be prepared by the user.
Reference:
A signal is not output for Alarm numbers 01,09,22,23, or 26.