Platinous User’s Manual
71
Operation
To power the lamp (equipment with 2A or more power)
Be sure to fit the line with a suitable capacity contactor and fuse.
FIGURE 6.42.
T
IME
SIGNAL
TERMINAL
(2A
OR
MORE
)
2. Turn the time signal ON for Steps 2 and 4 of the program.
Example 2: Following is an example 5-step program that charges specimens in steps 2 and 4.
FIGURE 6.43.
E
XAMPLE
PROGRAM
Procedure:
1. Connect the specimens to the time signal terminals and specimen power supply control terminals as
shown in the following diagrams.
Lam p
Fuse
Fuse
Power
supply
Time s ignal inlet
Target
temperature
(°C)
80
50
20
Time signal
ON
OFF
Step No.
1
2
3
4
5
Содержание Platinous ESL Series
Страница 1: ...PLATINOUS CHAMBER SERIES USER S MANUAL...
Страница 2: ......
Страница 6: ...iv Platinous User s Manual...
Страница 22: ...Precautions in Usage 16 Platinous User s Manual...
Страница 31: ...Platinous User s Manual 25 Names and Functions of Parts...
Страница 40: ...Names and Functions of Parts 34 Platinous User s Manual...
Страница 58: ...Installation 52 Platinous User s Manual...
Страница 80: ...Operation 74 Platinous User s Manual...
Страница 106: ...Troubleshooting 100 Platinous User s Manual...
Страница 107: ...Platinous User s Manual Chapter 9 Appendix...