Platinous User’s Manual
51
Installation
FIGURE 5.29.
W
ATER
LEVEL
CHECK
IN
WICK
PAN
Reference:
If water level in the wick pan is low or high, regulate it as follows:
1. Stop the supply of water and drain the water from inside the pan adjusting the water
clamp on the lower part of the water level regulator.
2. Remove the box nut on the water supply side.
3. Loosen the nut on the water supply side.
4. Adjust the water level to the standard line by turning the valve seat. Turn clockwise
when the water level is higher than the standard line and turn counter clockwise when
lower.
5. Tighten the nut, taking care so that the valve seat will not turn together with it.
6. Re-supply water after tightening the box nut and adjusting the water clamp and check
the water level described in Item #1 above.
7. Repeat steps 1-6 when the water level cannot be adjusted in one try.
Box Nut
Nut
Valve Seat
Water Clamp