Precautions in Usage
* Can be turned ON/OFF
** In some cases, refrigeration is shut OFF instead of capacity being reduced.
Absolute high/low limit temperature (& humidity) alarms and upper deviation limit temperature
alarm.
The absolute high/low limit alarms are completely independent of the target temperature and humidity.
They do not change when target temperature or humidity are changed.
Conversely, the upper deviation limit alarm is relative to the target temperature. If the target temperature is
changed, the deviation alarm temperature changes in proportion.
TABLE 2.2.
H
ARDWARE
/
SOFTWARE
SAFETY
DEVICES
Safety Device
Trip Point
Equipment
Response when
Tripped
Remarks
Overheat Protector
- Set approximately 20°C
higher than target tempera-
ture.
- Set below specimen’s max-
imum allowed temperature.
- Alarm displayed
- Buzzer sounded
- Heater shut OFF
- Humidifier shut OFF
(Temperature only
chambers excluded)
- Refrigerator shut OFF
- Air circulator shut OFF
- Specimen power sup-
ply control terminal
interrupted
N/A
Built into te
mp
erature &
h
umidi
ty controll
er
Temperature
Alarms
Absolute
High Limit
- Set approximately 15°C
higher than target tempera-
ture.
- Set below specimen’s max-
imum allowed temperature
- Max. 160°C
Absolute
Low Limit
- Set above following low
limit temperatures:
ESL/ESU: -40°C
ESX/ESZ: -75°C
- Set a minimum 5°C lower
than target temperature
- Set above specimen’s mini-
mum allowed temperature.
Upper
Deviation
Limit
- Set approximately 10°C
higher than target tempera-
ture.
- Alarm displayed
- Buzzer sounded*
- Heater shut OFF
When
tempera-
ture &
humidity
returns
within
range lim-
its, alarms
are auto-
matically
cleared.
Humidity
Alarms (Not
included on
temperature
only cham-
bers)
Absolute
High Limit
- Set approximately 10% RH
higher than target humidity.
- Set below specimen’s max-
imum allowed humidity.
- Alarm displayed
- Buzzer sounded*
- Humidifier shut OFF
Absolute
Low Limit
- Set approximately 10% RH
lower than target humidity.
- Set above specimen’s mini-
mum allowed humidity.
- Alarm displayed
- Buzzer sounded*
- Humidifier shut OFF
- Refrigeration capacity
lowered**
Содержание Platinous ESL Series
Страница 1: ...PLATINOUS CHAMBER SERIES USER S MANUAL...
Страница 2: ......
Страница 6: ...iv Platinous User s Manual...
Страница 22: ...Precautions in Usage 16 Platinous User s Manual...
Страница 31: ...Platinous User s Manual 25 Names and Functions of Parts...
Страница 40: ...Names and Functions of Parts 34 Platinous User s Manual...
Страница 58: ...Installation 52 Platinous User s Manual...
Страница 80: ...Operation 74 Platinous User s Manual...
Страница 106: ...Troubleshooting 100 Platinous User s Manual...
Страница 107: ...Platinous User s Manual Chapter 9 Appendix...