ets-lindgren.com
EMControl Command Set
85
*STB?
Description:
Query Status Byte. Prepares the controller to respond to the
contents of the Status Byte Register when queried.
The value returned is a decimal in the range of 0 to 255. This
value, when expressed in binary form, represents the eight-bit
value of the Status Byte Register.
Devices:
Turntables, towers
Syntax:
*STB?
Response:
XXX
: Integer value representing the bits of the Status Byte
Register. The bit pattern is as follows:
0
– Device Dependent Error (DDE)
: Set when the logical
AND of the Device Dependent Error Register and the Device
Dependent Error Enable Register is non-zero.
1
– Undefined
2
– Undefined
3
– Undefined
4
– Message Available (MAV)
: Set when there is data in the
device output queue waiting to be read over the GPIB.
5
– Event Status Bit (ESB)
: Set when the logical AND of the
Event Status Register and the Event Status Enable Register is
non-zero.
6
– Master Summary Status (MSS)
: Set when the logical
AND of the remaining bits of the Status Byte Register and the
Status Byte Enable Register is non-zero.
7
– Undefined
Example:
Write
sdSTB?\n
: Query status byte
Read response
Response is
16\n
: Message Available
See Also:
ERR?
,
*SRE
,
*SRE?
Содержание ETS-Lindgren EMControl 7006-001
Страница 1: ...Model 7006 001 EMControl Positioner Controller Plug In Card User Manual...
Страница 8: ...viii ets lindgren com This page intentionally left blank...
Страница 12: ...12 Introduction ets lindgren com This page intentionally left blank...
Страница 16: ...16 Specifications ets lindgren com This page intentionally left blank...
Страница 18: ...18 EMControl Plug In Card Installation ets lindgren com This page intentionally left blank...
Страница 88: ...88 EMControl Command Set ets lindgren com This page intentionally left blank...
Страница 90: ...90 Warranty ets lindgren com This page intentionally left blank...