ets-lindgren.com
EMControl Command Set
51
LV
Description:
Changes the lower limit of the device for vertical polarity. The
specified value must be less than the upper limit for the vertical
polarization and between
–999 and 999.
Devices:
Towers
Syntax:
LV [+/
–] XXX
+
: Optional for positive values
–
: Necessary for negative values
XXX
: Decimal value for the lower limit to be set in
centimeters. Leading zeroes are optional.
Response:
OK
, or an error (error number)
Example:
Write
sdLV 95\n
: Set vertical polarity lower limit to 95 cm
See Also:
LV?
LV?
Description:
Queries the lower limit of the device for vertical polarity.
The value returned is in XXX format depending upon the current
numeric mode. Negative values are preceded by a
– (minus sign).
Devices:
Towers
Syntax:
LV?
Response:
[
–]XXX
: Value of the lower limit for the vertical polarization
in centimeters
Example:
Write
sdLV?\n
: Query vertical lower limit
Read response
Response is
95\n
: Vertical lower limit is 95 cm
See Also:
LV
Содержание ETS-Lindgren EMControl 7006-001
Страница 1: ...Model 7006 001 EMControl Positioner Controller Plug In Card User Manual...
Страница 8: ...viii ets lindgren com This page intentionally left blank...
Страница 12: ...12 Introduction ets lindgren com This page intentionally left blank...
Страница 16: ...16 Specifications ets lindgren com This page intentionally left blank...
Страница 18: ...18 EMControl Plug In Card Installation ets lindgren com This page intentionally left blank...
Страница 88: ...88 EMControl Command Set ets lindgren com This page intentionally left blank...
Страница 90: ...90 Warranty ets lindgren com This page intentionally left blank...