490725-05
5
Product Safety Markings
Attention: See the instructions for use
Consult the
instructions for use
DO NOT use
tap water
Mains input
fuse rating
Potentially
hot surfaces
Product Information Markings
Manufacturer
Date of
manufacture
Serial number
Catalogue
number
The UKCA marking of
the product certifies
that it complies with
the Medical Devices
Regulations 2002 (SI
618), as amended
by the EU Exit
Regulations 2019 (SI
791) and 2020 (SI
1478)
2 Introduction & installation
2.1 Introduction
The Eschmann SES 2010 is an ‘N’ type, non-vacuum
autoclave with an 11 litre chamber. The autoclave runs
a 134°C cycle
with selectable drying.
The autoclave is intended for the sterilisation of solid,
unwrapped instruments and has been designed in
accordance with applicable standards.
2.2 Installation
We recommend that the autoclave is installed and
commissioned by an Eschmann or Eschmann Trained
Engineer. Contact Eschmann for installation, User
Training and Warranty Registration.
Note: This manual does not include configuration
settings to be made only by an Eschmann or
Eschmann Trained Engineer
2.3 Unpacking
Note: Recycle all packaging in accordance with
local recycling practices.
Remove the outer packaging and all packaging pieces.
CAUTION
Always move or lift the autoclave using at least two
people. Observe good manual handling techniques
at all times, with hands placed under the base on
each side of the autoclave.
Check for the following items (some are packed inside
the chamber - remove these):
User Manual Mains power cable
Drain tube
Tray carrier and tray lifter
Please note that contents may vary
2.4 Autoclave positioning
• Place the autoclave on a flat, level surface that is
both water and heat resistant.
• Ensure that the mains plug and power switch will
be easily accessible during use.
• There must be a clearance of at least 50mm to
the sides and rear of the unit and free space
above for filling the water reservoir.
• Clearance to the front to open the door: 300mm
• Ensure that the door is able to open fully.
• For autoclave environmental requirements, refer
to Appendix 5.
2.5 Mains connection
Connect the autoclave to a 230V AC mains power
socket using the supplied power cable.
Use only the Eschmann supplied mains power
cable. It is fitted with the correct fuse
.
Eschmann recommend the use of a suitable RCD
(Residual Current Device).
2.6 Product/ packaging Markings
o
r
0086
WEEE
recycling: Do
not dispose of
product at end
of life with other
waste