490725-05
29
Appendix 1 : Daily/weekly testing
Manufacturer’s testing must be carried out as outlined
below.
We recommend a daily Automatic Control Test, and the
completion of a Daily/Weekly Log Book (available from
Eschmann - 119857 Log Book (N Type).
All tests must be completed successfully.
Daily Testing
Automatic Control Test (with TST Indicator)
1. Load the autoclave and the TST Indicator as per the
manufacturer’s instructions.
2. Run the 134°C N type cycle.
3. Remove the TST Indicator at the end of the cycle.
4. Compare the TST Indicator with the Device’s
instructions.
WARNING!
If errors are detected when running test
cycles, or if the TST Indicator indicates a
failure, the cause of the failure MUST be
corrected and the test must be repeated
successfully before the autoclave is used to
process loads.
CAUTION
The TST Indicator must only be used and
stored as detailed in the Instructions For Use
supplied with the Device. Failure to do so
could lead to failure of the Device and
dangerously misleading results.
This device satisfies the requirement to perform a daily
steam penetration test as given in:
HTM01-05
Decontamination in primary care dental
practices.
EN 17665-1
Sterilisation of health care products. Moist
heat. Requirements for the development, validiation
and routine control of a sterilisation process for medical
devices.
DB 2002(06)
Device Bulletin, Benchtop Steam
Sterilisers - Guidance on purchase, Operation and
Maintenance, Medical Devices Agency, clause 5.3.1.
HTM01-01 (HTM2010/DB2002(06))
testing
If required, further Daily, Weekly, Quarterly and Annual
Testing may be necessary to satisfy this guidance.
Contact Eschmann for details on the automatic control
test.
Fig. 13. TST Indicators (Class 6)
To order:
REF 8793547
TST Indicators (Class 6) - 250 uses
Routine processing
A TST Indicator is to be placed into the centre of the
load for every cycle processing instruments.