490725-05
11
1
Stop the cycle
SES 2010
09:00
Cycle Type
Cycle No
Temp °C
Pressure kPa
Press and hold for 2 seconds to stop
the cycle.
2
The cycle has stopped
SES 2010
09:00
USER
STOPPED CYCLE
Load Not
Sterilised
Advisory Code:
2
Press this to
display details of
the cycle.
Refer to Section 11 for
further details.
Take care!
The load, chamber and
door will be VERY HOT.
WARNING!
Press and hold for 2 seconds to stop the cycle.
The load is NOT sterilised.
When a sterilising cycle is aborted the load must be treated
as non-sterile. The load must be sterilised by running the cycle again.
8 How to stop the cycle
Red cross
indicates cycle
failed.
There may be a delay
whilst the chamber
stabilises, allowing the
door to be opened.
5
SES 2010
09:00
Cycle Type
Cycle #
CYCLE
COMPLETE
Use Load
Immediately
This screen is only
displayed if drying was
selected, but the door
was not opened onto the
catch.
Unload autoclave following Section 7,
steps 2 and 3
Green tick indicates
cycle passed
If door NOT opened for drying
4
Drying Completed
SES 2010
09:00
DRYING
COMPLETE
Use Load
Immediately
Cycle Type
Cycle #
Open the door by releasing the catch. Unload
autoclave following Section 7,
step 3.
Press to view the
completed cycle log.
You will be prompted
to enter a PIN code.
Refer to Section 3.2.
Press to
accept cycle.
You will be prompted
to enter a PIN code.
Refer to Section 3.2.
Press to
accept cycle.
Press OK to
unlock door,
when OK displayed.
You will be prompted
to enter a PIN code.
Refer to Section 3.2.
Содержание Little Sister SES 2010