Appendix 4: Trouble shooting
Trouble shooting table
Fault
Action
Autoclave will not power-
up
Check the mains supply, mains connections, autoclave fuses and mains cable fuse. If
fault cannot be found, contact Eschmann.
Autoclave powers-up but
the screen is blank
Power autoclave off. Remove any item plugged into the USB port (Fig. 1, pg 4).
Power autoclave on. If unsuccessful, contact Eschmann.
Advisory
screens
Advisory
code
Description
Action
POWER
FAILURE
1
Power Failure
Check for intermittent mains supply and mains
connections. Run cycle again.
USER
STOPPED
CYCLE
2
User Abort
Run the required cycle.
SYSTEM
STOPPED
3
Door Lock Fail
Check for obstructions. Run cycle again.
11
Temperature Synchronisation Fail
Power autoclave off / on. Run cycle again.
15
Arbiters Disagree
Power autoclave off / on. Run cycle again.
16
Invalid Cycle Type
Power autoclave off / on. Run cycle again.
19
Temperature and Pressure Disagree
Visually check door seal. Clean mating face.
Run cycle again.
20
Temperature Sensors Disagree
Power autoclave off / on. Check chamber and
load is dry. Run cycle again.
21
T1 High During Sterilisation
Power autoclave off / on. Run cycle again.
23
T1 Low During Sterilisation
Power autoclave off / on. Run cycle again.
25
Not Enough Water in Reservoir
Fill with fresh water. Run cycle again.
26
Too Much Water in Reservoir
Empty waste water. Run cycle again.
33
Door Not Closed
Power autoclave off / on. Run cycle again.
34
Door Lock Fail
Power autoclave off / on. Run cycle again.
36
PR and/or GI Not In Running
Power autoclave off / on. Run cycle again.
41
Control System Error Occurred
Power autoclave off / on. Run cycle again.
42
Error Wrong Hold Time
Power autoclave off / on. Run cycle again.
The autoclave continually checks that it is able to run a
cycle and sterilise the load to the required standard.
If it encounters a problem it will inform you via the
Advisory Screens:
• Water Management Screens.
• Service Screens.
• Advisory Screens:
• POWER FAILURE
• USER STOPPED CYCLE
• SYSTEM STOPPED
• ERROR
Refer to Section 11 Advisory Screens
All Advisory Screens are identified internally using
an Advisory Code. These are shown in the Advisory
Screens Table below.
Most of the problems referred to by the Advisory
Screens can be rectified, allowing the cycle to
continue.
In most cases,
press OK to clear the Advisory
Screen
and follow the advice in the table below.
Note: The advisory screens that cannot be
cleared by the operator are the ERROR
screens. Contact Eschmann. Eschmann
require the Advisory Code from these
screens to determine the problem. See end
of table 1.
Advisory code table
Table 1: Advisory Codes