490725-05
9.1 Emptying the water reservoir
WARNING!
Water in the reservoir
can be very hot, take care when draining to
avoid contact with the hot water.
The water reservoir drain point is located as shown
below:
Remove the drain plug from the water reservoir
connector by pressing the metal thumb release on the
connector side.
Place the cut end of the drain tube into the sink, or a
container on a surface below the level of the autoclave
(Capacity: 5 litres or more).
Connect the drain tube to the water connector by
pressing the metal thumb release on the connector
side. This action opens the valve. The red drain tube clip
is now holding the water back.
Open the drain tube clip to drain the water.
9.2 Switch off
Switch off the autoclave at the mains switch (Fig 1, pg
4) when not in use. Leave it with the door ajar and the
chamber empty.
9.3 Cleaning
At the end of each day carry out the Cleaning and Care
instructions. Refer to Appendix 3.
We recommend that the water reservoir is
emptied, as a minimum,
at the end of each day.
9 After use
Push to open
Press to close
If using a container, remember to empty it. A five litre
container is sufficient to empty reservoir.
Once drained, release the drain tube from the
connector by pressing the thumb release on the side of
the connector. Refit the drain plug.
Check that the drain tube clip is not damaged or
leaking. If so,
replace as soon as possible
- refer to
Appendix 5: “Technical data”.