490725-05
15
Displayed when the autoclave experiences an Error.
1. When it appears, press the OK button to
acknowledge the warning and unlock the
door.
2. The door locked symbol at the top of the
screen will change to the door unlocked
symbol when the door is ready to be
opened to retrieve instruments.
The autoclave is prevented from being used as
this Error screen cannot be cleared.
Press to display the details required for
contacting Eschmann.
3. Switch the autoclave off and unplug from
the electrical wall socket.
Eschmann use only (PIN protected).
WARNING
Cycle
Stopped
SES 2010
09:00
CRITICAL
ERROR
SYSTEM
STOPPED
Load Not
Sterilised
Advisory Code:
999
Displayed when the autoclave prevents operation
during a cycle
.
Press to display the details required for
contacting Eschmann.
1. Note the Advisory Code. When it appears,
press OK to acknowledge the warning and
clear the screen.
2. Refer to Appendix 4 and complete the
Actions against the relevant Advisory
Code.
3. Run the cycle again. If the warning
persists, contact Eschmann.
Eschmann use only (PIN protected).
SYSTEM STOPPED
ERROR
All errors require you to contact Eschmann.
SES 2010
09:00
ERROR
Contact
Eschmann
Advisory Code:
999
SES 2010
09:00
ERROR
Contact
Eschmann
Advisory Code:
999
WARNING
Cycle
Stopped
SES 2010
09:00
CRITICAL
ERROR
SYSTEM
STOPPED
Advisory Code:
999
Displayed when the autoclave prevents operation when
a
cycle is not in progress
.
Press to display the details required for
contacting Eschmann.
1. Note the Advisory Code. When it appears,
press OK to acknowledge the warning and
clear the screen.
2. Refer to Appendix 4 and complete the
Actions against the relevant Advisory
Code.
3. Run the cycle again. If the warning
persists, contact Eschmann.
Eschmann use only (PIN protected).
SYSTEM STOPPED
The screen changes to: