Ingenuity in
infection control
since 1830
User Manual
Little Sister
SES 2010
Non-vacuum
Autoclave
Страница 1: ...Powered by TCPDF www tcpdf org Ingenuity in infection control since 1830 User Manual Little Sister SES 2010 Non vacuum Autoclave...
Страница 2: ...on 5 2 3 Unpacking 5 2 4 Autoclave positioning 5 2 5 Mains connection 5 2 6 Product packaging Markings 5 3 Controls 6 3 1 Touch screen menu operation 6 3 2 The Keypad 6 3 3 The Progress Screen 7 3 4 A...
Страница 3: ...shooting In this case DO NOT switch off power until discharge of steam into the reservoir has stopped Do not attempt to open the chamber door until the unit has cooled 1 3 Limitations of use THIS IS A...
Страница 4: ...on Bar CYCLE SELECTION Menu SETTINGS Menu RECORDS Menu 1 Chamber door 2 Door handle 3 Door catch 4 Touch screen display Fig 2 5 Mains on off switch 6 RS232 port 7 USB host port for USB memory stick 8...
Страница 5: ...ote Recycle all packaging in accordance with local recycling practices Remove the outer packaging and all packaging pieces CAUTION Always move or lift the autoclave using at least two people Observe g...
Страница 6: ...ever requires a User PIN Cycle Delay Keypad SES 2010 09 00 2 3 4 5 6 7 8 9 0 1 3 Controls 3 1 Touch screen menu operation Touch screen The autoclave must be switched on to use the touch screen The au...
Страница 7: ...and maintenance of the autoclave Maintaining a record of training attendance including evidence of trainee understanding Refer to Section 13 Security to set up security PINS and choose a security lev...
Страница 8: ...truments to be sterilised Load Cycle type Solid unwrapped instruments intended for immediate use N Maximum autoclave load 4 5kg Load trays using the supplied tray lifter All instruments MUST be thorou...
Страница 9: ...essure kPa Indicates drying not selected WARNING During operation Do not remove or cover the water reservoir lid Fig 1 pg 4 Caution hot steam is vented Cycle progress Cycle Chart Cycle type Load type...
Страница 10: ...door when OK button displayed WITHOUT DRYING WITH DRYING 1 Sterilisation complete SES 2010 09 00 Cycle Type Cycle No Temp C Pressure kPa 2 Positioning door for drying Take care The load chamber and do...
Страница 11: ...cle failed There may be a delay whilst the chamber stabilises allowing the door to be opened 5 SES 2010 09 00 Cycle Type Cycle CYCLE COMPLETE Use Load Immediately This screen is only displayed if dryi...
Страница 12: ...ve The red drain tube clip is now holding the water back Open the drain tube clip to drain the water 9 2 Switch off Switch off the autoclave at the mains switch Fig 1 pg 4 when not in use Leave it wit...
Страница 13: ...nd decontamination when disposing of the door seal 3 Update the Settings Menu counter for the door seal Updating the Settings Menu It is important to update the door seal counter in the Maintenance Me...
Страница 14: ...available whilst the autoclave is depressurising in preparation for the door being opened This may take up to two minutes at which point one of the following Warning screens will be displayed Refer to...
Страница 15: ...schmann 1 Note the Advisory Code When it appears press OK to acknowledge the warning and clear the screen 2 Refer to Appendix 4 and complete the Actions against the relevant Advisory Code 3 Run the cy...
Страница 16: ...Start screens Press to view the Cycle Selection menu The 134 N cycle is the factory default for each new cycle The current cycle type is shown on the Start screen 2 Cycle Select SES 2010 09 00 Cycle S...
Страница 17: ...s to accept the delay time shown and go to 5 below or Press to enter a new delay time Press to return without change 4 Cycle Delay Keypad SES 2010 09 00 2 3 4 5 6 7 8 9 0 1 00 00 Insert the Delay Time...
Страница 18: ...the 24 hour clock Press to return without change 3 Date Time Date entry SES 2010 09 00 Date Time 13 00 01 10 2019 Daylight Saving Set the autoclave s date and time Press OK to accept Press to return...
Страница 19: ...larms Cycle Alarms Cycle Keypress Note that alarms are always sounded The selection highlighted in blue is the current selection PIN management SES 2010 09 00 PIN Management 0000 3 2 1 Manager Setting...
Страница 20: ...highlighted setting Press OK to accept Press to return without change Replace the filter and or seal before resetting the counter s When you reset an item on this screen the item is marked with an arr...
Страница 21: ...ince Last Download will download all records produced since the last time it was used to download records Press to return without change Selecting All Since Last Download will begin the download immed...
Страница 22: ...atted memory stick is full memory stick was unplugged search criteria returned no data Correct and try again By date 1 SES 2010 09 00 Set FROM TO Dates High Security Today s Date 01 30 __ __ __ Set FR...
Страница 23: ...codes will allow you to know Who started a cycle High security level Who stopped a cycle All security levels Who unlocked the door at the end of a cycle All security levels Who cleared Service screens...
Страница 24: ...a Manager PIN 0000 is an unused PIN is a used PIN 4 Pin Entry SES 2010 09 00 PIN Configuration User 2 0002 2 Cycle Delay Keypad SES 2010 09 00 2 3 4 5 6 7 8 9 0 1 0002 Select to Edit PIN or status lev...
Страница 25: ...ue area of the screen Press to accept Press to return without change 3 SES 2010 09 00 Security Level Low Security Medium Security High Security Select required security level Use the up down buttons t...
Страница 26: ...older on your PC desktop To create the back up folder right click on the desktop and select New then Folder Type in the name of the new folder We suggest SES 2010 If you are backing up more than one u...
Страница 27: ...elow 2 Click the arrow to select your printer 3 Click to print the file Printing options toggle to display all the data in the cycle or only major events toggle to select double sided or single sided...
Страница 28: ...hh mm ss dd mm yy ERR error STS status at cycle end PCV pressure change value Service and certification tags SVC service expired date cycle count of service expiry CRT certification expired date of ce...
Страница 29: ...ator must only be used and stored as detailed in the Instructions For Use supplied with the Device Failure to do so could lead to failure of the Device and dangerously misleading results This device s...
Страница 30: ...pendix 5 Maximum Loads If in doubt about loads and processing details contact Eschmann Unwrapped solid items only The autoclave is ONLY suitable for unwrapped solid items Refer to Section 1 3 Limitati...
Страница 31: ...ogenic material from instruments used in the Health sector Eschmann are able to advise on how to recycle and dispose of the product correctly Example 1 6 x HFiT10 Cassettes 60 Instruments Example 2 2...
Страница 32: ...al legislation WARNING Should the autoclave ever leak disconnect from the mains electrical supply drain the reservoir and call Eschmann Autoclave fuses Fuse x 2 Fig 1 pg 4 Refer to Appendix 5 Technica...
Страница 33: ...ion Power autoclave off on Run cycle again 25 Not Enough Water in Reservoir Fill with fresh water Run cycle again 26 Too Much Water in Reservoir Empty waste water Run cycle again 33 Door Not Closed Po...
Страница 34: ...cycle again 175 Pressure Not Equalised Power autoclave off on Run cycle again 176 Safety Relay Fail Visually check door seal Clean mating face Run cycle again or Leave autoclave to cool Safety relay...
Страница 35: ...Mismatch Contact Eschmann ERROR 18 Pressure Switch Failed Contact Eschmann 177 Failed To Update Calibration Record Contact Eschmann 178 Calibration Invalid Contact Eschmann 179 No valid configuration...
Страница 36: ...4 5 kg WATER AND RESERVOIR Water Reservoir capacity Use per cycle 2500 ml 600ml Proportion re cycled Temperature of drained water flow rate 81 C 1 2 l min Heat transmitted by autoclave in one hour 72...
Страница 37: ...et Pressure 2 85 bar Hydrostatic Test Pressure 4 15 bar Pollution degree 2 SAFETY FEATURES Eschmann autoclaves are designed to be safe and effective No changes should be made except by an Eschmann Eng...
Страница 38: ...lid load Small porous loads Full porous load Narrow Lumen Dryness solid load Dryness porous load Non condensable gases Load chart Cycle Data Sterilisation Temperature 134 137 C Sterilisation Pressure...
Страница 39: ...nd cycle codes Absolute Pressure Cycle stages 1 to 5 Detailed N type cycle information Progress Indicator Cycle stage Stage Detail 1 2 1 Filling and 2 heating to sterilisation temperature 3 Sterilisin...
Страница 40: ...FiT Cassettes HFiT 10 Examination Stainless Steel Cassette 13 5cm x 18cm 8722812 Stainless Steel Blue 8722813 Stainless Steel Red 8722814 Stainless Steel Green 8722815 Stainless Steel Yellow 8722816 S...
Страница 41: ...Powered by TCPDF www tcpdf org 490725 05 41...
Страница 42: ...n this publication may be subject to change without prior notice Service The Autoclave must only be serviced by an Eschmann or Eschmann trained engineer We also recommend a program of ongoing service...