3.4.4.2 Procedure
Zero
checking:
With the [Sample] or [Span] key, select the inlet used and wait for measurement to stabilize.
The analyzer response should be between -1 and +1 ppb.
Next carry out a zero reference cycle by pressing the [Z. Ref.] key of the
« Cycle »
submenu in
the
« MEASUREMENT »
menu.
Calibration
correction:
Automatic:
Select the measurement range corresponding to the span gas titration, as well as the inlet
used and press the [Auto] key (for use of this function, refer to section 0
« MEASUREMENT
Instantaneous »
). The analyzer automatically modifies its span coefficient according to the
programmed titration. The recommended duration for the auto-calibration is 1200 seconds.
NOTE :
During the cycle, the displayed NO
2
measurement uses the previous calibration coefficient.
This calibration coefficient is updated when the Auto-calibration cycle is finished.
If the new calibration coefficient is out of the limits of the previous calibration coefficient
50 %, the
analyzer will send a calibration alarm message. In this case, verify the following:
Fluid circuit of calibration
Concentration generated by the used gas source
Concentration of the programmed span gas
Programmed gas inlet
If one of the items above is wrong, correct it and carry out a new auto-calibration cycle. If they are all
correct, carry out a manual adjustment of the calibration coefficient as follows:
Manual
adjustment:
Select the inlet where the gas is connected by pressing the [Sample] or [Span] key. Wait for
the measurement to stabilize. Program the new calibration coefficient in the
« SPAN
Coefficients»
screen.
Calculation of the new calibration coefficient
:
x
=
offset)
(without
value
read
value
gas
span
K
K
(old)
new
K
(new)
is the new calibration coefficient
K
(old)
is the previous calibration coefficient
ATTENTION :
It is advised to write down the « K
(old)
» before modifying it because it will be
cancelled by the validation of the «K
(new)
».
3.4.4.3
Use of Auto-calibration cycle
To program this cycle, refer to section 3.3.3.3 –
« SPAN
Cycles »
menu.
To configure the concentration, refer to section 3.3.3.2 –
« SPAN
Select gas
»
menu.
The gas generation system will be permanently connected to the span gas inlet of the analyzer. The
recommended timing for the automatic calibration is 1200 seconds.
Содержание AS32M
Страница 8: ...EnvironnementS A AS32M Duplication prohibited MAY 2013 0 8 Page intentionally left blank...
Страница 26: ...EnvironnementS A AS32M Duplication prohibited AUGUST 2013 2 2 Page intentionally left blank...
Страница 38: ...EnvironnementS A AS32M Duplication prohibited DECEMBER 2013 2 14 Page intentionally left blank...
Страница 104: ...EnvironnementS A AS32M Duplication prohibited DECEMBER 2013 3 66 Page intentionally left blank...
Страница 106: ...EnvironnementS A AS32M Duplication prohibited AUGUST 2013 4 2 Page intentionally left blank...
Страница 111: ...Duplication prohibited AS32M EnvironnementS A AUGUST 2013 4 7 Page intentionally left blank...
Страница 126: ...EnvironnementS A AS32M Duplication prohibited AUGUST 2013 5 2 Page intentionally left blank...
Страница 133: ...Duplication prohibited AS32M EnvironnementS A DECEMBER 2013 5 9 Page intentionally left blank...
Страница 134: ...EnvironnementS A AS32M Duplication prohibited DECEMBER 2013 5 10 Figure 5 1 Module board of the AS32M...
Страница 140: ...EnvironnementS A AS32M Duplication prohibited DECEMBER 2013 5 16 Page intentionally left blank...
Страница 142: ...EnvironnementS A AS32M Duplication prohibited AUGUST 2013 6 2 Page intentionally left blank...
Страница 160: ...EnvironnementS A ESTEL Board Duplication prohibited JUNE 2009 18 Page intentionally left blank...
Страница 168: ...EnvironnementS A DNP ARM7 board Duplication prohibited NOVEMBER 2009 8 Page intentionally left blank...
Страница 186: ...EnvironnementS A USB Flash Drive Duplication prohibited NOVEMBER 2012 14 Page intentionally left blank...