6 Basic System Setup
VIEW 370/373
28
(PC-000855-MA Issue 6, December 2021)
ENVEA
UK Ltd
1.
Select the
Setup | More Settings | Add a new device
.
2.
Choose the correct
Channel Type
from the list and set the following
parameters:
a.
Modbus Address
: set to the address set on the sensor DIP switches.
b.
Group and Device Name
: set to identify the channel (default: Stack1
Dust).
Figure 12: More Settings
–
Sensor settings screen.
3.
Select
Save
to apply the settings and return to the main menu.
For some sensor types, this will initiate a download to the sensor. If this
fails, an error message is displayed, and the display will revert to the sensor
settings screen. Recheck the
Modbus Address
and retry. If the download
fails repeatedly, press
Cancel
on page 38 to
diagnose the problem.
For sensors that do not need to download configuration settings, a sensor
channel will be added automatically to the Bar Graph/Text screen. Open the
Bar Graph
screen and check the sensor is reading correctly. If the sensor
fails to
communicate for 10 seconds, a ‘No Response’ alarm will appear for
that device. If this happens, select
Edit an Existing Device,
and check the
Modbus Address
. If the alarm remains, refer to
38 to diagnose the problem.
6.9.2
Editing Sensor/Device Settings
1.
Select
Setup
|
More Settings | Edit an existing device
.
2.
Select the
Device
to be edited and change the settings as required.
3.
Select
Save
to apply the new settings to the sensor/device.
6.9.3
Deleting a Sensor/Device
1.
Select
Setup
|
More Settings | Delete an existing device
.
2.
Select the
Device
to be deleted, and then press ENTER to confirm.