![envea VIEW 370 Скачать руководство пользователя страница 13](http://html1.mh-extra.com/html/envea/view-370/view-370_user-manual_2415518013.webp)
VIEW 370/373
2 Technical Data
ENVEA
UK Ltd
(PC-000855-MA Issue 6, December 2021)
5
2
Technical Data
2.1
Process and Application Conditions
Stack temperature ranges
A
-20 to +250 °C (-4 to 482 °F)
Option: up to 400 (752 °F) / 800 °C (1,472 °F)
Stack diameter
Ø 100 mm to 6 m (4
”
to 20 ft., nom.)
Application conditions
Suitable for measurement in non-condensing flue gases.
Note
: not suitable for use downstream of ESPs or in
applications with water droplets.
Sensor
–
Key data
Measurement capability
0
–
10 to 0
–
500 mg/m³
(automatic range changes)
Long-term zero drift
<0.1 mg/m³
Velocity dependency
For use in processes with flow above the 8
–
20 m/s
range with no restrictions.
Constant velocity is required outside this range.
Inspection frequency
Every 6 months
Environmental information
Temperature range
-20 to +50 °C (-4 to 122 °F), ambient
2.2
Sensor Specification
VIEW 370 / 373 Sensor
Sensor variants
Standard sensor:
Insulated sensor:
Passive/active sensor:
High-temperature sensor:
0
–
100 °C / 0
–
250 °C
0
–
250 °C
0
–
250 °C / 0
–
400 °C
0-400 °C / 0
–
800 °C
Sensor body material
Standard:
316 Stainless steel
Passive/active:
316 Stainless steel
Fully insulated:
PTFE-coated 316 Stainless steel
Weight
1.8 kg (4 lb), standard sensor
Stack connection
1½
″
BSP (female)
B
Enclosure
Dimensions
W 122 x H 122 x D 80 mm (5 x 5 x 3.2
”
)
Protection rating
IP65
Enclosure material
Die-cast aluminium (epoxy coated)
Cable entries
2x M16 cable glands (one blanked
ex-works
)
A
At monitoring point. Please note that imperial temperatures given are nominal values.
B
Ensure the opening/hole in stack wall is at least Ø45 mm.